YSI 9300 and 9500 Photometers User Manual ®

Página 2 - Direct-Read Photometers

7  Set the dilution factor to ‘On’ or ‘Off’. If the dilution factor option is set to ‘On’, the instrument will allow the entry of a numerical fact

Página 3 - Y- PT 282

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ Oxygen is a major

Página 4

Test Procedure 1 Filter sample if necessary to obtain a clear solution. 2 Fill the test tube with sample to the 10 ml mark. 3 Add one Sulfitest N

Página 5

PHOTOMETER TEST INSTRUCTIONS ________________________________________________________________________________________ Zinc compounds are

Página 6

SEPARATION OF RESIDUALS The photometer is programmed for both zinc and the copper correction procedure. Use program Phot 35 Zinc (+ Co

Página 7

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ The Alkalinity of

Página 8

Reagents and Equipment YSI Alkaphot M Tablets YSI Alkaphot P Tablets YSI 9300 or 9500 Photometer Round Test Tubes, 10 ml glass (PT 595) Test Proc

Página 9

NOTE The expression of alkalinity results sometimes causes confusion. It is normal practice to express the result as mg/l CaCO3 (calcium c

Página 10

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ Iron occurs wide

Página 11

Reagents and Equipment YSI Iron MR No 1 Tablets YSI Iron MR No 2 Tablets YSI Citrate IR Tablets YSI 9300 or 9500 Photometer Round Test Tubes, 10 ml

Página 12

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ Hydrazine is used a

Página 13

8 Units The photometer offers the choice of result expressed in mg/l, ppm, mmol/l, µmol and g/l. Sample Number (9500 Only) A unique number may be

Página 14

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ Formulations c

Página 15

Test Procedure 1 Filter sample if necessary to obtain a clear solution. 2 Fill test tube with sample to the 10 ml mark. 3 Add one Molybdate No 1

Página 16

PHOTOMETER TEST INSTRUCTIONS _________________________________________________________________________________________ Nitrites and nitrite

Página 17 - D ilutio n Facto r

Test Procedure 1 Filter sample if necessary to obtain a clear solution. 2 Using the measuring syringe take 1 ml of the sample. Transfer to the tes

Página 18 - Phot 007

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ The use of organop

Página 19 - Phot 008

Reagents and Equipment YSI Oxidising OP Tablets YSI OP-A Tablets YSI OP-B Tablets YSI OP-AX Tablets YSI 9300 or 9500 Photometer YSI Test Tube, 20 ml

Página 20 - T im e r

Test Procedure - Correction Factor If it is suspected that the sample contains orthophosphate, carry out the following correction procedu

Página 21

PHOTOMETER TEST INSTRUCTIONS _________________________________________________________________________________________ The YSI Chloridol te

Página 22

Reagents and Equipment YSI Acidifying CD Tablets YSI Chloridol Tablets YSI 9300 or 9500 Photometer Round Test Tubes,

Página 23

5 Select the appropriate program number on the photometer for the test range required. 6 Take the photometer reading in usual manner (see photomet

Página 24

9 Locking System Mode Settings (9500 Only) It is possible to 'lock’ the system settings so that these cannot be tampered with or altered a

Página 25

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ Pure water exhibits

Página 26 - TABLET REAGENT SYSTEM

Test Procedure 1 Filter sample through a GF/B filter paper. 2 Fill a test tube with filtered sample to the 10 ml mark. 3 Fill a test tube with de

Página 27

PHOTOMETER TEST INSTRUCTIONS ________________________________________________________________________________________ Turbidity is an impo

Página 28

PHOTOMETER TEST INSTRUCTIONS _________________________________________________________________________________________ The presence of d

Página 29

Test Procedure Read the Oxygen test instructions leaflet contained in the CHEMetrics Vacu-Vials pack. Observe these various recommendations rega

Página 30

PHOTOMETER TEST INSTRUCTIONS _________________________________________________________________________________________ DISSOLVED OXYGEN/

Página 31 - PHOTOMETER TEST INSTRUCTIONS

Test Procedure Read the Oxygen test instructions leaflet contained in the CHEMetrics Vacu-Vials pack. Observe these various recommendations rega

Página 32

PHOTOMETER TEST INSTRUCTIONS _________________________________________________________________________________________ PHMB Polyhexamethylb

Página 33

Test Procedure 1 Fill the test tube with sample to the 10 ml mark. 2 Add one PHMB-Phot tablet crush and mix to dissolve. 3 Select Phot 52 on phot

Página 34

PHOTOMETER TEST INSTRUCTIONS _________________________________________________________________________________________ Nickel does not occ

Página 35

10 USB (9500 Only) The USB interface allows communication between the instrument and a PC. There is a choice of two operating modes – Hard Drive an

Página 36

Test Procedure 1 Fill test tube with sample to the 10 ml mark. 2 Add one Nickeltest No 1 tablet, crush and mix to dissolve. Ensure tablet is compl

Página 37

PHOTOMETER TEST INSTRUCTIONS ___________________________________________________________________________________________ Phenols and substi

Página 38

Reagents and Equipment YSI Phenoltest No 1 Tablets YSI Phenoltest No 2 Tablets YSI Phenoltest PR Tablets YSI 9300 or 9500 Photometer Round Test

Página 39 - CHLORINE DIOXIDE

PHOTOMETER TEST INSTRUCTIONS _________________________________________________________________________________________ Chromium may be pre

Página 40

Reagents and Equipment For Hexavalent Chromium :- YSI Chromicol No 1 Tablets YSI Chromicol No 2 Tablets YSI 9300 or 9500 Photometer YSI Test Tube

Página 41

Test Procedure - Total Chromium 1 Fill the pretreatment tube with sample to the 20 ml mark. 2 Add one level spoonful of Chromicol CR powder

Página 42

PHOTOMETER TEST INSTRUCTIONS _________________________________________________________________________________________ Silicon, in the

Página 43

Test Procedure 1 Fill the test tube with sample to the 10 ml mark. 2 Add one Silica No 1 tablet, crush and mix to dissolve. 3 Add one Silica No 2

Página 44

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ Chlorine dioxide i

Página 45

Important Note Chlorine dioxide is extremely volatile and can be lost from solution very easily. Extreme care must be taken when extracting and disp

Página 46

11 4 INTERFACE CONNECTIONS AND DATA MEMORY (9500 ONLY) Stored data can be accessed by recall to the instrument display (see ‘View Log’). A

Página 47

PHOTOMETER TEST INSTRUCTIONS _________________________________________________________________________________________ Chlorine dioxide i

Página 48

Important Note Chlorine dioxide is extremely volatile and can be lost from solution very easily. Extreme care must be taken when extracting and disp

Página 49

Item # Y-PT 282 Rev A Drawing # Y-PT 282 © 2010 YSI IncorporatedV1-11/10

Página 51

12 5 TAKING PHOTOMETER READINGS The photometer is very simple to use. Screen prompts guide the user towards the test result. The follo

Página 52 - __________________________

13 The BLANK TUBE is a test tube filled only with the water being tested only. It is important to use the actual water to be tested to provide a tru

Página 53

14 If the sample number option is pre-selected, then the following display will appear, for example (9500 only): Enter or confirm the sample num

Página 54

15 3 The instrument will be set automatically. After a few seconds the following display will appear: Place SAMPLE TUBE in the test chamber

Página 55

16 Continuation Tests (Certain Tests Only) 1 Select ‘Follow-On’ and press [OK] while the result is displayed of the currently running

Página 56

YSI 9300 and 9500 Direct-Read Photometers User Manual YSI, Inc. 1725 Brannum Lane Yellow Springs, OH 45387 Tel: 800-897-4151 (+1 937-767-72

Página 57

17 Reading in Transmittance and Absorbance When taking readings in Transmittance or Absorbance mode, use the  and  keys to step through the

Página 58

18 To program user-defined calibrations: Turn the instrument ‘ON’, select ‘System’ menu and press [OK]. Scroll through the options to the USB entry

Página 59

19 An example is shown below: USER CALIBRATION 900 Chlorine mmol/l 500nm Cl2 0,0 0.174,0.50 0.481,1.50 0.733,2.50 0.854,3.00 1.022,4.00 1.086,4.50 1

Página 60

20 6 CARE AND MAINTENANCE The photometer is designed to give long and trouble-free operation. Care must be taken, however, to avoid

Página 61

21 Photometer Up-Grade (9500 only) It is now possible to upgrade the photometer with new test calibrations using a computer. This will ensure that u

Página 62

22 Limitation of Warranty This Warranty does not apply to any YSI product damage or failure caused by: 1 Failure to install, operate or use

Página 63

Table of Contents Photometer Instructions PHOT.INST. TABLET REAGENT SYSTEM Alkalinity, Total (Alkaphot)

Página 64

Direct-Reading Photometer Program Schedule TABLET REAGENT SYSTEM Instruction Sheet Number Reagent System Parameter Photometer Program Number - - Tr

Página 65

Instruction Sheet Number Reagent System Parameter Photometer Program Number PHOT.23. Nitrate (Nitratest) Nitrate Nitrogen Phot 23 PHOT.24. Nitrite

Página 66

PHOTOMETER TEST INSTRUCTIONS _________________________________________________________________________________________ Natural and treat

Página 67

INDEX 1 INTRODUCTION ... 1 Fe

Página 68 - Test Procedure

Test Procedure 1 Fill test tube with sample to the 10 ml mark. 2 Add one Alkaphot tablet, crush and mix until all of the particles have dissolved.

Página 69

PHOTOMETER TEST INSTRUCTIONS _________________________________________________________________________________________ Aluminum sulphate is widel

Página 70

Sample Collection Aluminum is readily absorbed on to the surfaces of sample containers, particularly glass containers. To avoid loss of aluminum, co

Página 71

PHOTOMETER TEST INSTRUCTIONS _________________________________________________________________________________________ Ammonia occurs as

Página 72

Test Instructions 1 Fill test tube with sample to the 10 ml mark. 2 Add one Ammonia No 1 tablet and one Ammonia No 2 tablet, crush and mix to dis

Página 73

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ Bromine and brom

Página 74

Reagents and Equipment YSI DPD No 1 Clear Tablets YSI DPD Nitrite Tablets YSI 9

Página 75

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ Chlorine and chlor

Página 76

Reagents and Equipment YSI DPD No 1 Tablets YSI DPD No 2 Tablets YSI DPD No 3 Tablets YSI 9300 or 9500 Photometer Round Test Tu

Página 77

PHOTOMETER TEST INSTRUCTIONS _________________________________________________________________________________________ CHLORINE DIOXIDE

Página 78

1 1 INTRODUCTION The YSI 9300 and 9500 direct-read photometers are designed to give long and trouble-free operation. To ensure the best resul

Página 79

Reagents and Equipment YSI DPD No 1 Tablets YSI DPD No 3 Tablets YSI DPD Glycine Tablets YSI DPD Acidifying Tablets YSI DPD Neutralising Tablets YSI

Página 80

PHOTOMETER TEST INSTRUCTIONS _________________________________________________________________________________________ Chlorine dioxide is

Página 81

Reagents and Equipment YSI DPD No 1 Tablets YSI DPD No 3 Tablets YSI DPD Glycine Tablets YSI DPD Acidifying Tablets YSI DPD Neutralizing Tables YSI

Página 82 - multiply result by 4.4

PHOTOMETER TEST INSTRUCTIONS _________________________________________________________________________________________ Chlorine and chlori

Página 83

Reagents and Equipment YSI DPD No 1 Tablets YSI DPD Oxystop Tablets (Optional) YSI DPD No 3 Tablets YSI 9300 or 9500 Photometer Round Test Tubes,

Página 84

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ Chlorine and chlorine relea

Página 85

Reagents and Equipment YSI DPD No 1 Tablets YSI DPD No 2 Tablets YSI DPD No 3 Tablets YSI 9300 or 9500 Photometer Round Test Tu

Página 86

PHOTOMETER TEST INSTRUCTIONS _________________________________________________________________________________________ CHLORINE DIOXIDE

Página 87

Reagents and Equipment YSI DPD No 1 Tablets YSI DPD No 3 Tablets YSI DPD Glycine Tablets YSI DPD Acidifying Tablets YSI DPD Neutralising Tablets YSI

Página 88

PHOTOMETER TEST INSTRUCTIONS _________________________________________________________________________________________ Chlorine dioxide i

Página 89

2 Instrument Layout Test and Number Keys Multi-Size Cell Holder OK (Enter) Key Scroll and Selection Keys On/Off Key USB Interfac

Página 90

Reagents and Equipment YSI DPD No 1 Tablets YSI DPD No 3 Tablets YSI DPD Glycine Tablets YSI DPD Acidifying Tablets YSI DPD Neutralizing Tables YSI

Página 91

PHOTOMETER TEST INSTRUCTIONS _________________________________________________________________________________________ Chlorine and chlorin

Página 92

Reagents and Equipment YSI DPD No 1 Tablets YSI DPD Oxystop Tablets (Optional) YSI DPD No 3 Tablets YSI 9300 or 9500 Photometer Round Test Tubes,

Página 93

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ Chlorine and chlor

Página 94

Test Instructions 1 Rinse test tube with sample leaving two or three drops of sample in the tube. 2 Add one DPD No 4 tablet, crush tablet and then

Página 95

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ Chlorine and chlo

Página 96

Test Procedure 1 Fill test tube with sample to the 10 ml mark. 2 Add one Acidifying GP tablet and one Chlorine HR tablet. Crush tablets and mix to

Página 97

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ Copper occurs natu

Página 98

Separation of Copper Residuals The direct-reading photometer is programmed for both free and total copper. Use program Phot 10 Free Copper, then

Página 99

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ Calcium hardness

Página 100 -

3 Features and Technical Specification Application For application in general water testing using Palintest tablet reagent systems and Palin

Página 101

Test Procedure 1 Filter sample if necessary to obtain a clear solution. 2 Fill the test tube with sample to the 10 ml mark. 3 Add one Calcicol No

Página 102

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ Cyanuric acid is

Página 103

Test Procedure 1 Fill test tube with sample to the 10 ml mark. 2 Add one Cyanuric Acid tablet and allow to disintegrate for at least t

Página 104

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ Fluoride occurs n

Página 105

TEST PROCEDURE 1 Fill test tube with sample to the 10 ml mark. 2 Add one Fluoride No 1 tablet, crush and mix to dissolve. 3 Add one Fluoride No 2

Página 106

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ Water hardness i

Página 107

Test Procedure 1 Filter sample if necessary to obtain a clear solution. 2 Fill test tube with sample to the 10 ml mark. 3 Add one Hardicol No 1 t

Página 108

PHOTOMETER TEST INSTRUCTIONS _________________________________________________________________________________________ Hydrogen peroxide is

Página 109

Test Procedure 1 Rinse test tube with sample leaving 2 to 3 drops of sample in the tube. 2 Add one Hydrogen Peroxide LR tablet, crush and then fi

Página 110

PHOTOMETER TEST INSTRUCTIONS _________________________________________________________________________________________ Hydrogen peroxide

Página 111

4 2 OPERATING PRINCIPLE The YSI photometers are instruments that measure color intensity. Light is passed through a test tube containing

Página 112

Test Procedure 1 Fill test tube with sample to the 10 ml mark. 2 Add one Acidifying PT tablet and one Hydrogen Peroxide HR tablet. Crush tablets

Página 113

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ Iron occurs widely

Página 114

Reagents and Equipment YSI Iron LR Tablets YSI 9300 or 9500 Photometer Round Test Tubes 10 ml glass (PT 595) Test Procedure 1 Fill the test tube w

Página 115

PHOTOMETER TEST INSTRUCTIONS _________________________________________________________________________________________ Iron occurs widely i

Página 116

Test Procedure 1 Fill test tube with sample to the 10 ml mark. 2 Add one Iron HR tablet, crush and mix to dissolve. 3 Stand for one minute to all

Página 117

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ Manganese-containi

Página 118 - Range: 0 - 50,000 mg/l NaCl

Reagents and Equipment YSI Manganese No 1 Tablets YSI Manganese No 2 Tablets YSI 9300 or 9500 Photometer Round Test Tubes, 10 ml glass (PT 595)

Página 119 - Multiply by

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ Magnesium is a

Página 120

TEST PROCEDURE 1 Using the measuring syringe take a 1 ml sample of the water under test. Transfer to the round test tube and make up to t

Página 121

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ MOLYBDATE HR Formu

Página 122

5 Replacing the Batteries The battery compartment in the base of the instrument is secured by four screws. To replace the batteries, remove the cove

Página 123

Reagents and Equipment YSI Molybdate No 1 HR Tablets YSI Molybdate No 2 HR Tablets YSI 9300 or 9500 Photometer Round Test Tubes, 10 ml glass (PT

Página 124

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ Nitrates are norm

Página 125

Reagents and Equipment YSI Nitratest Powder (Spoon Pack) YSI Nitratest Tablets YSI Nitricol Tablets YSI Nitr

Página 126

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ Nitrites are foun

Página 127

Test Procedure 1 Fill round test tube with sample to the 10 ml mark. 2 Add one Nitricol tablet, crush and mix to dissolve. 3 Stand for 10 minutes

Página 128 - ______________________

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ Ozone is used for

Página 129

Reagents and Equipment YSI DPD No 4 Clear Tablets YSI DPD Glycine Tablets YSI 9300 or 9500 Photometer Round Test Tubes, 10 ml glass (PT 595) Se

Página 130

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ pH measurement i

Página 131

Notes 1 The color range of the phenol red test is yellow, through orange, to red. The formation of an intense purple coloration shows that the indi

Página 132

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ Phosphates are ex

Página 133

6 3 GENERAL PHOTOMETER OPERATION The photometer is controlled by a simple intuitive menu system:  The highlight indicates the active line or se

Página 134 - For Hexavalent Chromium :

Reagents and Equipment YSI Phosphate No 1 LR Tablets YSI Phosphate No 2 LR Tablets YSI 9300 or 9500 Photometer Round Test Tubes, 10 ml glass (PT

Página 135

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ Phosphates are ext

Página 136

Test Procedure 1 Fill test tube with sample to the 10 ml mark. IN THE CASE OF SAMPLES CONTAINING SILICA (>20 mg/l SiO2) ONLY :- 2 Add one Phosp

Página 137

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ Potassium is an abu

Página 138

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ Silicon, in the

Página 139

Test Instructions 1 Fill test tube with sample to the 10 ml mark 2 Add one Silica No 1 tablet, crush and mix to dissolve. Stand for five minutes t

Página 140

PHOTOMETER TEST INSTRUCTIONS __________________________________________________________________________________________ Sulfates occur natu

Página 141

Reagents and Equipment YSI Sulfate Turb Tablets YSI 9300 or 9500 Photometer Round Test Tubes, 10 ml glass (PT 595) Test Procedure 1 Fill test tube

Página 142 - V1-11/10

PHOTOMETER TEST INSTRUCTIONS _________________________________________________________________________________________ Natural waters co

Página 143

Sample Collection To prevent loss of sulphide collect the sample carefully with a minimum of agitation or aeration. Test the sample as s