aD0!
a (service request response)
“2” is number values in “D0” resp:
apu or aputu
“p” is ± pressure value
“u” is + units indicator. Use “XUP”
command to set units. If “XUM=1”
command is executed then ±
temperature value and units code (tu)
also appended.
M1
Start non concurrent
factory pressure
measurement
(factory calibrated value
always in psi)
aM1!
:
aD0!
attt1
:
a (service request response)
“1” is number values in “D0” resp:
ap
“p” is ± pressure value (psi)
M2
Start non concurrent
temperature
measurement
aM2!
:
aD0!
attt2
:
a (service request response)
“2” is number values in “D0” resp:
atu
“t” is signed temperature, “u” is = units:
0-Celsius, 1=Fahrenheit. Use “XUT”
command to set units.
M3
Start non concurrent
scale factor
measurement
(user slope, user offset,
and field calibration
offset)
aM3!
:
aD0!
a0013
:
a (service request response)
“3” is number values in “D0” resp:
asoc
“s” is signed user slope
“o” is signed user offset (psi)
“c” is signed field cal. Offset
M4
Start non concurrent
standards lab scale
factor measurement
(standards lab
calibration slope and
calibration offset)
aM4!
:
aD0!
a0012
:
a (service request response)
“3” is number values in “D0” resp:
aso
“s” is signed cal. Slope
“o” is signed cal. offset
M5
Start non concurrent
high accuracy PCB
temperature
aM5!
:
aD0!
attt2
:
a (Service Request response)
“2” is number values in “D0” resp:
atu
“t” is signed temperature, “u” is = units:
0-Celsius, 1=Fahrenheit. Use “XUT”
command to set units is si
M6
Start non concurrent
battery voltage
measurement
aM6!
:
aD0!
a0021:
:
a (Service request response)
“01” is number values in “D0” resp:
av
“v” is the battery voltage
M7
Start non concurrent
pressure and
temperature
measurement
aM7!
:
aD0!
attt4
4 is number values in “D0” resp:
aputu
p is ± pressure value, first u is = units
index. Units set by “XUP” command. t
is ± temperature value, second u is +0 =
Celsius or +1 = Fahrenheit. Use “XUT”
command to set units.
YSI Incorporated Data Scout Advanced 101