USER MANUALEnglish, Français, Español, PortuguêsPro1020User ManualDOCUMENT #605187
1213When Quick DO Cal is enabled, press and hold the calibration key for 3 seconds while in the Run screen. Next, highlight Dissolved Oxygen and pre
1415Selecting a Dissolved Oxygen Membrane:Membrane TypeFlow Dependence After 4 MinutesRequired Sample MovementTypical Response Time (T-95)5913, 5908 Y
1617mVs can help you determine if you are performing a good calibration and the condition of the pH electrode. When ORP is enabled in the ISE Sensor T
1819Parameter Reset DefaultsISE Sensor Type pHISE UnitspHISE Auto Stable Off (0.0 % Change and 10 seconds)pH Buffer SetUSATemperature Units °CPressure
2021If using a Galvanic sensor, wait approximately 5 to 10 minutes for the cup to become completely saturated. The instrument’s Auto Shutoff should be
22237. If the calibration is unsuccessful, an error message will display on the screen. Press the Cal key to exit the calibration error message and r
24259. Rinse the sensor and place it in the third pH buffer (4/4.01 or 10/9.18). If necessary, use the up and down arrow keys to adjust the pH buffer
2627temperature, dissolved oxygen and two conductivity parameters. Each data point is referenced with a data set number, 01 through 50.SAVING DATAFrom
2829EXITING DATA MODEWhile in Data mode, press enter to access the bottom functions. Next, highlight the ESC-Exit box and press enter to return to the
3031SENSOR MAINTENANCE - TEMPERATUREYou must keep the temperature sensor free of build up. No additional maintenance is required. A toothbrush can be
Item #605187Rev A, January 2014For the latest version of this manual,visit ysi.com©2014 YSI Incorporated.The YSI logo is a registered trademarks of YS
3233Remove the protective red cap by pulling it straight off the sensor. Or, unscrew and remove the used cap membrane by holding the sensor while unsc
3435Silver AnodeAfter extended use, a thick layer of Silver Chloride (AgCl) builds up on the silver anode (figure 11) reducing the sensitivity of the
3637bottle that was originally included with the sensor. This can be filled with buffer 4 solution. If you no longer have the storage bottle, simply p
3839Symptom Possible SolutionDO readings are inaccurate.1. Verify correct DO sensor/membrane type selection in the System Setup menu.2. Verify user-
4041Part Number Description605202 Galvanic DO Sensor605203 Polarographic Sensor603059 Flow cell, 203 ml 603056 Flow cell mounting spike603075 Carrying
4243RECYCLINGYSI is committed to reducing the environmental footprint in the course of doing business. Even though materials reduction is the ultimate
Item # 605187Rev A January 2014©2014 YSI Incorporated.
iAuto Shutoff ...17Resetting the System Setup Menu to Factory Default ...17C
ii1THIS PAGE LEFT INTENTIONALLY BLANKINTRODUCTIONThank you for purchasing the YSI Pro1020, an instrument from the YSI Professional Series product fam
23Figure 1. Pro1020 with battery cover removed. Notice battery symbols indicating polarities.The waterproof instrument case is sealed at the factory a
45iWhen a port is empty on the cable, the bulkhead connector is not water-proof. Do not submerge the cable without a sensor installed! Submerging the
67CONNECTING THE PROBE/CABLE ASSEMBLY TO THE INSTRUMENTTo connect the cable, align the keys on the cable connector to the slots on the instrument conn
89 NAVIGATING THE RUN SCREENWhen in the Run screen, the up and down arrow keys will move the highlighted box along the bottom options. Once a box
1011EXITING THE SYSTEM SETUP MENUTo exit the System Setup menu, press the down arrow key until the ESC - Exit box is highlighted, then press enter to